The Journal of Arthroplasty
Volume 18, Issue 3 , Pages 361-366, April 2003

Observations on retrieved glenoid components☆☆

Department of Orthopaedic Surgery, Inselspital, University of Berne, Switzerland

Received 14 September 2001; accepted 5 September 2002.

Abstract 

To obtain more information on the pattern of damage of prosthetic glenoid components, we analyzed 7 retrieved glenoid components. The consecutive series included 2 standard polyethylene components and 5 highly crystalline polyethylene glenoids (Hylamer; DePuy Dupont Orthopaedics, Warsaw, IN) retrieved 3 to 12 years after implantation. At revision, 4 of 5 Hylamer components were fractured. Common wear patterns were i) deformation and crumbling of the rim, particularly at the inferior hemicircumference, probably caused by direct contact of the humerus with the prosthetic component; ii) roughening (abrasion and scratching) of the adjacent articulating surface; and iii) concentric and congruous wear centered posteriorly. Available glenoid components may cover an excessive sector of the head. This can result in mechanical restriction of glenohumeral motion and abutment of the humerus against the glenoid rim. Abutment may cause major shear forces and therefore cause glenoid loosening. The value of articular surface mismatch is questionable because retrieved glenoids were worn to a conforming joint. © 2003 Elsevier Inc. All rights reserved.

Keywords:  shoulder, arthroplasty, glenoid component, complication, radial mismatch, glenohumeral impingement

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 No benefits or funds were received in support of this study.

☆☆ Reprint requests: Ralph Hertel, MD, Department of Orthopedic Surgery, Inselspital, University of Berne, 3010 Berne, Switzerland.

 0883-5403/03/1803-0019$30.00/0

PII: S0883-5403(02)06249-6

doi:10.1054/arth.2003.50048