Five Cases of Failure of the Tibial Polyethylene Insert Locking Mechanism in One Design of Constrained Knee Arthroplasty

Received 8 March 2010; accepted 18 July 2010. published online 01 September 2010.

Abstract 

We describe 5 cases of failure of the locking mechanism of the polyethylene insert and tibial base-plate in one design of constrained condylar knee prosthesis due to disengagement of the locking screw. Loosening of the screw is believed to occur because of a counterclockwise torque created by the axial rotation of the femur on the tibia that occurs as the knee extends during gait. This torque is transmitted via the highly rotationally constrained femoral housing and tibial post to the locking screw. These failures suggest that an alternative locking mechanism should be considered for this prosthesis.

Keywords: knee arthroplasty, constrained condylar prosthesis, locking screw, failure

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 No benefits or funds were received in support of the study.

PII: S0883-5403(10)00413-4

doi:10.1016/j.arth.2010.07.013

The Journal of Arthroplasty
Volume 26, Issue 6 , Pages 976.e21-976.e24, September 2011