    • Corresponding Author InformationReprint requests: Taek Rim Yoon, MD, Center for Joint Disease and Brain Korea 21 Project for Biomedical Human, Resources Center at Chonnam National University, Chonnam National University Hwasun Hospital, 160, Ilsim-Ri, Hwasun-Eup, Hwasun-Gun, Jeonnam 519-809, Korea.

Received 4 August 2010; accepted 6 October 2011. published online 24 November 2011.

Abstract 

We retrospectively reviewed records of 1002 patients who underwent a third-generation ceramic-on-ceramic total hip arthroplasty. Fifteen patients complained of squeaking, at any time after their arthroplasty. Of these 15 patients, 14 were evaluated clinically and radiologically. These 14 patients were found to have significantly higher mean body mass index of 25.8 kg/m2 as compared with all the other patients who did not complain of squeaking (mean, 23.7 kg/m2) (P = .022). The acetabular opening angle was significantly lower (mean, 34°) than a matched control group (mean, 38°) (P = .016). Limb length shortening of more than 5 mm was observed in 12 (85.7%) of the 14 patients as compared with only 4 (28.6%) of 14 patients in the matched control group. Flexion and sitting cross legged were identified as the movements, which most commonly (11/12) resulted in squeaking. The incidence of squeaking was found to be low (1.5%), and we identified high body mass index, decreased acetabular opening angle, and limb length shortening as factors associated with squeaking.

Keywords: ceramic-on-ceramic, squeaking, total hip arthroplasty

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 The Conflict of Interest statement associated with this article can be found at doi:10.1016/j.arth.2011.10.001.

PII: S0883-5403(11)00522-5

doi:10.1016/j.arth.2011.10.001

The Journal of Arthroplasty
Volume 27, Issue 6 , Pages 909-915, June 2012