Reliability of a PowerPoint Method for Wear Measurement After Total Hip Arthroplasty

A Retrieval Study Using 3-Dimensional Laser Scanning
Published:April 13, 2012DOI:https://doi.org/10.1016/j.arth.2011.12.025

      Abstract

      The purpose of this study was to evaluate the reliability of the PowerPoint (PP) (2007 Version; Microsoft, Redmond, Wash) method for measuring polyethylene liner wear after total hip arthroplasty. Seventeen retrieved polyethylene liners were included in this study. Wear volumes were calculated using the PP, the Dorr and Wan, and 3-dimensional (3D) laser scanning methods. Spearman correlation coefficients for wear volume results indicated strong correlations between the PP and 3D laser scanning methods (range, 0.89-0.93). On the other hand, Spearman correlation analysis revealed only moderate correlations between the Dorr and Wan and 3D laser scanning methods (range, 0.67-0.77). The PP method can be used to monitor linear wear after total hip arthroplasty and could serve as an alternative method when computerized methods are not available.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to The Journal of Arthroplasty
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Callaghan J.J.
        • Pedersen D.R.
        • Olejniczak J.P.
        • et al.
        Radiographic measurement of wear in 5 cohorts of patients observed for 5 to 22 years.
        Clin Orthop Relat Res. 1995; 317: 14
        • Charnley J.
        • Halley D.K.
        Rate of wear in total hip replacement.
        Clin Orthop Relat Res. 1975; 112: 170
        • Livermore J.
        • Ilstrup D.
        • Morrey B.
        Effect of femoral head size on wear of the polyethylene acetabular component.
        J Bone Joint Surg Am. 1990; 72: 518
        • Dorr L.D.
        • Wan Z.
        Ten years of experience with porous acetabular components for revision surgery.
        Clin Orthop Relat Res. 1995; 319: 191
        • Pollock D.
        • Sychterz C.J.
        • Engh C.A.
        A clinically practical method of manually assessing polyethylene liner thickness.
        J Bone Joint Surg Am. 2001; 83: 1803
        • Yamaguchi M.
        • Bauer T.W.
        • Hashimoto Y.
        Three-dimensional analysis of multiple wear vectors in retrieved acetabular cups.
        J Bone Joint Surg Am. 1997; 79: 1539
        • Shaver S.M.
        • Brown T.D.
        • Hillis S.L.
        • et al.
        Digital edge-detection measurement of polyethylene wear after total hip arthroplasty.
        J Bone Joint Surg Am. 1997; 79: 690
        • Glyn-Jones S.
        • Isaac S.
        • Hauptfleisch J.
        • et al.
        Does highly cross-linked polyethylene wear less than conventional polyethylene in total hip arthroplasty? A double-blind, randomized, and controlled trial using roentgen stereophotogrammetric analysis.
        J Arthroplasty. 2008; 23: 337
        • Bragdon C.R.
        • Kwon Y.M.
        • Geller J.A.
        • et al.
        Minimum 6-year followup of highly cross-linked polyethylene in THA.
        Clin Orthop Relat Res. 2007; 465: 122
        • Vandenbussche E.
        • Saffarini M.
        • Hansen U.
        • et al.
        Measurement of femoral head penetration in polyethylene using a 3-dimensional CT-scan technique.
        Acta Orthop. 2010; 81: 563
        • Jedenmalm A.
        • Noz M.E.
        • Olivecrona H.
        • et al.
        A new approach for assessment of wear in metal-backed acetabular cups using computed tomography: a phantom study with retrievals.
        Acta Orthop. 2008; 79: 218
        • Barrack R.L.
        • Lavernia C.
        • Szuszczewicz E.S.
        • et al.
        Radiographic wear measurements in a cementless metal-backed modular cobalt-chromium acetabular component.
        J Arthroplasty. 2001; 16: 820
        • Ebramzadeh E.
        • Sangiorgio S.N.
        • Lattuada F.
        • et al.
        Accuracy of measurement of polyethylene wear with use of radiographs of total hip replacements.
        J Bone Joint Surg Am. 2003; 2378: 85
        • Sychterz C.J.
        • Young A.M.
        • Engh C.A.
        Effect of radiographic quality on computer-assisted head penetration measurements.
        Clin Orthop Relat Res. 2001; 386: 150
        • Dorr L.D.
        • Wan Z.
        Comparative results of a distal modular sleeve, circumferential coating, and stiffness relief using the Anatomic Porous Replacement II.
        J Arthroplasty. 1996; 11: 419
        • Devane P.A.
        • Bourne R.B.
        • Rorabeck C.H.
        • et al.
        Measurement of polyethylene wear in metal-backed acetabular cups. I. Three-dimensional technique.
        Clin Orthop Relat Res. 1995; 319: 303
        • Martell J.M.
        • Berdia S.
        Determination of polyethylene wear in total hip replacements with use of digital radiographs.
        J Bone Joint Surg Am. 1997; 79: 1635
        • Bowden A.E.
        • Kurtz S.M.
        • Edidin A.A.
        Validation of a micro-CT technique for measuring volumetric wear in retrieved acetabular liners.
        J Biomed Mater Res B Appl Biomater. 2005; 75: 205
        • Raimondi M.T.
        • Sassi R.
        • Pietrabissa R.
        A method for the evaluation of the change in volume of retrieved acetabular cups.
        Proc Inst Mech Eng H. 2000; 214: 577
        • Jedenmalm A.
        • Nilsson F.
        • Noz M.E.
        • et al.
        Validation of a 3D CT method for measurement of linear wear of acetabular cups.
        Acta Orthop. 2011; 82: 35
        • Gallo J.
        • Havranek V.
        • Cechova I.
        • et al.
        Wear measurement of retrieved polyethylene ABG 1 cups by universal-type measuring microscope and x-ray methods.
        Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2006; 150: 321
        • Chuter G.S.
        • Cloke D.J.
        • Mahomed A.
        • et al.
        Wear analysis of failed acetabular polyethylene: a comparison of analytical methods.
        J Bone Joint Surg Br. 2007; 89: 273
        • Walker P.S.
        • Blunn G.W.
        • Lilley P.A.
        Wear testing of materials and surfaces for total knee replacement.
        J Biomed Mater Res. 1996; 33: 159
        • Jasty M.
        • Goetz D.D.
        • Bragdon C.R.
        • et al.
        Wear of polyethylene acetabular components in total hip arthroplasty. An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations.
        J Bone Joint Surg Am. 1997; 79: 349
        • Kabo J.M.
        • Gebhard J.S.
        • Loren G.
        • et al.
        In vivo wear of polyethylene acetabular components.
        J Bone Joint Surg Br. 1993; 75: 254
        • Thomsen M.
        • Burandt C.
        • Gortz A.
        • et al.
        An optical method for determining deformation and form changes in polyethylene surfaces of explanted acetabular cups of hip endoprostheses.
        Biomed Tech (Berl). 1999; 44: 247
        • Yun H.H.
        • Jajodia N.K.
        • Myung J.S.
        • et al.
        Use of slide presentation software as a tool to measure hip arthroplasty wear.
        J Arthroplasty. 2009; 24: 1210
        • Callaghan J.J.
        • Dysart S.H.
        • Savory C.G.
        The uncemented porous-coated anatomic total hip prosthesis. Two-year results of a prospective consecutive series.
        J Bone Joint Surg Am. 1988; 70: 337
        • Jacobs C.A.
        • Christensen C.P.
        • Greenwald A.S.
        • et al.
        Clinical performance of highly cross-linked polyethylenes in total hip arthroplasty.
        J Bone Joint Surg Am. 2007; 2779: 89
        • Thibault S.
        • Borra E.F.
        Telecentric three-dimensional sensor with a liquid mirror for large-object inspection.
        Appl Opt. 1999; 38: 5962
        • Derbyshire B.
        The estimation of acetabular cup wear volume from two-dimensional measurements: a comprehensive analysis.
        Proc Inst Mech Eng H. 1998; 212: 281
        • Isaac G.H.
        • Dowson D.
        • Wroblewski B.M.
        An investigation into the origins of time-dependent variation in penetration rates with Charnley acetabular cups—wear, creep or degradation?.
        Proc Inst Mech Eng H. 1996; 210: 209
        • Bould M.
        • Barnard S.
        • Learmonth I.D.
        • et al.
        Digital image analysis: improving accuracy and reproducibility of radiographic measurement.
        Clin Biomech (Bristol, Avon). 1999; 14: 434
        • Digas G.
        • Karrholm J.
        • Thanner J.
        • et al.
        Highly cross-linked polyethylene in cemented THA: randomized study of 61 hips.
        Clin Orthop Relat Res. 2003; 417: 126
        • Moore K.D.
        • Barrack R.L.
        • Sychterz C.J.
        • et al.
        The effect of weight-bearing on the radiographic measurement of the position of the femoral head after total hip arthroplasty.
        J Bone Joint Surg Am. 2000; 82: 62
        • Digas G.
        • Karrholm J.
        • Thanner J.
        • et al.
        The Otto Aufranc Award. Highly cross-linked polyethylene in total hip arthroplasty: randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis.
        Clin Orthop Relat Res. 2004; 429: 6
        • Saikko V.
        • Calonius O.
        • Keranen J.
        Wear of conventional and cross-linked ultra-high-molecular-weight polyethylene acetabular cups against polished and roughened CoCr femoral heads in a biaxial hip simulator.
        J Biomed Mater Res. 2002; 63: 848
        • Wan Z.
        • Boutary M.
        • Dorr L.D.
        Precision and limitation of measuring two-dimensional wear on clinical radiographs.
        Clin Orthop Relat Res. 2006; 449: 267
        • Collier M.B.
        • Kraay M.J.
        • Rimnac C.M.
        • et al.
        Evaluation of contemporary software methods used to quantify polyethylene wear after total hip arthroplasty.
        J Bone Joint Surg Am. 2003; 2410: 85